Sunteți pe pagina 1din 13

Proiect nr.

12-110/2008

Microsenzori magnetici implantabili pentru aplica ii medicale

MEDISENS

Coordonator: Institutul National de Cercetare-Dezvoltare pentru Fizica Tehnica IFT Iasi Director proiect: CS II Dr. Tibor-Adrian vri

INCDFT-IFT Iasi - MEDISENS

Date initiale

Parteneri in consortiu:
CO: Institutul National de Cercetare-Dezvoltare pentru Fizica Tehnica IFT Iasi P1: Universitatea Tehnica Gh. Asachi Iasi Facultatea de Electrotehnica Facultatea P2: Universitatea de Medicina si Farmacie Gr. T. Popa Iasi de Bioinginerie Medicala Partener CO P1 P2 TOTAL Buget (lei) 1.450.000 250.000 250.000 1.950.000

Cofinantare (lei) 0 0 0 0

Adresa web a proiectului: http://www.phys-iasi.ro/MEDISENS.htm Adresa este functionala


INCDFT-IFT Iasi - MEDISENS
2

Date initiale

Obiectivul/rezultatul final al proiectului:


Obiectivul principal este realizarea unor prototipuri de microsenzori ce utilizeaza microfirele magnetice nanostructurate ca elemente sensibile pentru aplicatii medicale: (1) retele de microfire pentru detectia semnalelor biomagnetice extrem de mici, cu aplicatii in magneto-cardiografia la temperatura ambianta; (2) microsenzori pentru detectia si monitorizarea de la distanta a tensiunilor mecanice si a temperaturii din implanturi si organe.

Elemente de noutate ale proiectului; originalitate:


Realizarea de microfire magnetice biocompatibile cu pastrarea caracteristicilor utile pentru aplicatii in microsenzori; Utilizarea efectului magneto-impedantei gigant in aplicatii medicale; Dezvoltarea unei metode de detectie de la distanta a tensiunilor mecanice / deformarilor din implanturi.

Beneficiarii rezultatelor; potentialul de aplicare in economie:


Beneficiarii rezultatelor sunt partenerii din consortiu. Drepturile de proprietate intelectuala vor fi protejate prin brevete de inventie. Anticipam un interes crescut al companiilor cu activitati in domeniul senzorilor medicali si biosenzorilor (de exemplu SIEMENS), fata de principiile de functionare si de prototipurile de microsenzori.
INCDFT-IFT Iasi - MEDISENS
3

Stadiul de realizare al proiectului in 2010: Etapele 1 si 2 (2009)


Etapa 1:

Bugetul de stat: 300.000 lei (CO) Cofinantare: 0 Activitati efectuate: Realizarea tehnologiei de preparare a microfirelor amorfe precursoare. Achizitii realizate: Camera de ecranare electromagnetica Notebook 1 buc 1 buc (finantare partiala);

Etapa 2:

Bugetul de stat: 23.567 lei (CO) Cofinantare: 0 Activitati efectuate: Experimentari privind tratamentul termic in vederea stabilirii parametrilor optimi (timp, temperatura, atmosfera) pentru obtinerea starii nanocristaline. Achizitii realizate: nu s-au facut.
INCDFT-IFT Iasi - MEDISENS
4

Stadiul de realizare al proiectului in 2010: Etapa 3 (2010)


Etapa 3:

Bugetul de stat: 223.919 lei, din care: CO: 200.000 lei P2 (UMF): 23.919 lei

Cofinantare: 0 Activitati efectuate: Experimentari privind functionalizarea microfirelor magnetice nanostructurate ca materiale biocompatibile; Studiul proprietatilor magnetice a microfirelor nanostructurate in regim static, respectiv dinamic; Studiul efectului magneto-impedantei gigant in microfire magnetice nanostructurate; Studiul polarizarii electrice dependente de structura magnetica (antena inteligenta) in diferite conditii de tensionare mecanica sau deformare.

Achizitii realizate: Anexa baie de organ - Dispozitiv pentru masurarea impedantei electrice (P2).
INCDFT-IFT Iasi - MEDISENS
5

Implicarea tinerilor cercetatori

La activitatile realizate in cadrul proiectului a participat un numar important de tineri cercetatori (varsta sub 35 ani), dupa cum urmeaza: Etapa 1 2 3 Total persoane 20 4 14 Din care, tineri: 8 (40%) 1 (25%) 5 (35.7%)

Trei tineri cercetatori si-au sustinut tezele de doctorat in domenii conexe celui abordat in proiect. Activitatea desfasurata in cadrul proiectului a contribuit semnificativ la formarea lor ca cercetatori si la finalizarea tezelor lor de doctorat, avand titlurile:
Sorin Corodeanu Materiale magnetice compozite; Gabriel Ababei Studiul proprietatilor magnetice ale materialelor amorfe si nanostructurate in domeniul frecventelor inalte; Mihai Tibu Senzori si actuatori pe baza de materiale magnetice amorfe.

INCDFT-IFT Iasi - MEDISENS

Indicatori de rezultat

precursori pentru rezultatele finale

Am realizat functionalizarea microfirelor magnetice nanostructurate cu trei tipuri de polimeri biocompatibili. Am obtinut astfel solutia pentru elementul sensibil implantabil ce va fi utilizat la realizarea microsenzorilor implantabili pentru detectia si monitorizarea de la distanta a tensiunilor mecanice din implanturi si organe (precursor pentru rezultatul final). Testele efectuate indica faptul ca microfirele magnetice nanostructurate acoperite cu straturi de polimer biocompatibil sunt stabile la degradare in organismul uman. Am studiat proprietatile magnetice ale microfirelor magnetice nanostructurate in regim static si dinamic. Studiul rezultat demonstreaza superioritatea proprietatilor microfirelor magnetice nanostructurate realizate in etapele anterioare pentru aplicatiile medicale avute in vedere (precursor pentru rezultatul final). Am studiat efectul magneto-impedantei gigant in microfire magnetice nanostructurate. Studiul realizat evidentiaza o dependenta puternica a amplitudinii efectului in functie de tensiunile mecanice la care sunt supuse microfirele. Aceasta comportare, impreuna cu valorile foarte mari ale amplitudinii efectului (peste 600% in anumite cazuri) le recomanda pentru utilizarea in microsenzori magnetici pentru detectia semnalelor biomagnetice foarte mici (precursor pentru rezultatul final). Am stabilit principiul detectiei de la distanta pe baza interactiunii elementelor sensibile realizate din microfire magnetice nanostructurate cu un camp electromagnetic de frecventa ridicata (2,3 2,7 GHz), care este modulat prin intermediul fenomenului de magneto-impedanta de catre un camp de modulare, in diferite conditii de tensionare/deformare mecanica. Datele din studiul realizat arata ca principiul corespunde aplicatiei de monitorizare de la distanta a tensiunilor mecanice din implanturi. Metoda dezvoltata arata sensibilitatea microfirelor utilizate la intindere, torsiune si incovoiere, constituind un precursor pentru rezultatul final, sensibilitatea putand fi ajustata prin dimensiunile microfirelor magnetice selectate pentru rolul de element de detectie.

INCDFT-IFT Iasi - MEDISENS

Indicatori de rezultat

Articole:
1. H. Chiriac, M. Tibu, T.-A. vri, Domain wall propagation in nanocrystalline glass-coated microwires , IEEE Transactions on Magnetics, Vol. 45, No. 10, 2009, pp. 4286-4289; 2. S. Corodeanu, T.-A. vri, N. Lupu, H. Chiriac, Magnetization process and GMI effect in as-cast nanocrystalline microwires , IEEE Transactions on Magnetics, Vol. 46, No. 2, 2010, pp. 380-382; 3. H. Chiriac, M. Lostun, T.-A. vri, Surface magnetization processes in amorphous microwires , IEEE Transactions on Magnetics, Vol. 46, No. 2, 2010, pp. 383-386; 4. T.-A. vri, M. Grigoras, H. Chiriac, Magnetoresistance effect in soft magnetic amorphous microwires , Journal of Applied Physics, Vol. 107, 09A317 (2010); 5. H. Chiriac, S. Corodeanu, G. Ababei, T.-A. vri, Magnetic behavior of rapidly quenched submicron amorphous wires , Journal of Applied Physics, Vol. 107, 09A301 (2010); 6. V. Balan, M.I. Popa , D. Hritcu, L. Verestiuc, Preparation and characterization of magnetic particles based on magnetite and chitosan , ISSN 2066-7817, pages 322-327, 2010; 7. Capitol carte: Biodegradable polymers for controlled release of drugs L. Verestiuc, E.C. Tanase, in Medical Applications of Polymers, Editors: M. Popa, R. M. Ottenbritte, C.V. Uglea, American Scientific Publishers, USA, 2010.

Cerere inregistrare brevet:


Sistem de ecranare selectiva a radiatiei electromagnetice de inalta frecventa nr. depozit la OSIM: a 2010 01173/25.11.2010, autori: H. Chiriac, T.-A. vri, G. Ababei.

Model functional:
Instalatie de preparare a microfirelor amorfe precursoare cu dimensiuni si proprietati controlate (pe baza tehnologiei realizate in etapa 1 din 2009).

INCDFT-IFT Iasi - MEDISENS

Activitatile prevazute a se realiza in 2011

In planul de realizare al proiectului este prevazut un numar de 16 activitati pentru ultimul an al proiectului. Acestea au ramas, fiind practic mutate din primii doi ani ai proiectului, datorita problemelor de finantare sumele primite nu au coincis cu cele din planul de realizare initial, ritmicitatea lucrarilor fiind data peste cap. Astfel, bugetul ramas pentru 2011 din bugetul initial total al proiectului este de 1.402.514 lei, reprezentand 72% din total. In conditiile in care suma pentru 2011 nu poate depasi bugetul acceptat initial pentru 2011, respectiv 700.000 lei, bugetul pentru 2011 al proiectului va fi injumatatit, respectiv redus cu suma de 702.514 lei, ceea ce inseamna renuntarea la o parte din activitatile prevazute pentru 2011. Vom pastra acele activitati care ne permit continuarea muncii incepute si realizarea in cea mai mare masura a microsenzorilor propusi initial in proiect. Astfel, vom realiza: Prototipul microsenzorului neimplantabil alcatuit dintr-o retea de microfire pentru detectia semnalelor biomagnetice de valori extrem de mici; Prototipul microsenzorului pentru detectia si monitorizarea la distanta a tensiunilor mecanice din implanturi; Experimentarea si verificarea prototipurilor de microsenzori realizati.

INCDFT-IFT Iasi - MEDISENS

Activitatile prevazute a se realiza in 2011

Ne vom concentra doar pe activitatile care conduc strict la, si sunt indispensabile pentru realizarea microsenzorilor medicali propusi. Obiectivul principal al proiectului se mentine. In aceste conditii, elementele de originalitate se pastreaza in totalitatea lor. Rezultatele preconizate pentru 2011 sunt prototipurile microsenzorilor (2 produse), articole ISI (3) si inca o cerere de brevet. Suma de 700.000 lei este necesara pentru realizarea activitatilor selectate pentru anul 2011.

INCDFT-IFT Iasi - MEDISENS

10

Activitati 2011 comparativ cu cele prevazute in planul de realizare


Activitati prevazute in planul de realizare pentru 2011
4.1 Studiul efectului diferitelor conditii de presiune sau temperatura asupra polarizarii electrice dependente de structura magnetica 4.2 Studiul metodelor de optimizare a sensibilitatii microsenzorilor propusi la actiunea diversilor factori externi 4.3 Studiul zgomotului magnetic intrinsec in microfire magnetice nanostructurate si realizarea sistemelor hibride formate din microfire si din circuite cu zgomot extrem de redus si a unor sisteme de masura la distanta pentru microsenzorii propusi 4.4 Stabilirea parametrilor tehnologiei de obtinere a microfirelor magnetice nanostructurate si optimizarea acesteia in vederea obtinerii de materiale cu dimensiuni, microstructura si proprietati controlate pentru realizarea prototipurilor de microsenzori 4.5 Studiul structural al microfirelor magnetice nanostructurate si corelarea cu proprietatile magnetice 4.6 Realizarea tehnologiei de functionalizare a microfirelor magnetice nanostructurate ca materiale biocompatibile 4.7 Studiul stabilitatii structurale, mecanice si magnetice a microfirelor magnetice nanostructurate in urma procesarii acestora pentru realizarea sistemelor hibride 4.8 Proiectarea si realizarea prototipului de microsenzor neimplantabil alcatuit dintr-o retea de microfire pentru detectia semnalelor biomagnetice de valori extrem de mici 4.9 Realizarea partii electronice a prototipului microsenzorului pentru detectia semnalelor biomagnetice extrem de mici 4.10 Realizarea unei camere ecranate magnetic in vederea experimentarii prototipului de microsenzor pentru detectia semnalelor biomagnetice extrem de mici 4.11 Proiectarea si realizarea prototipurilor microsenzorilor pentru detectia si monitorizarea la distanta a tensiunilor mecanice si a temperaturii din implanturi si organe 4.12 Realizarea partii electronice a prototipurilor microsenzorilor pentru detectia si monitorizarea la distanta a tensiunilor mecanice si a temperaturii din implanturi si organe

INCDFT-IFT Iasi - MEDISENS

11

Activitati 2011 comparativ cu cele prevazute in planul de realizare

Activitati prevazute in planul de realizare pentru 2011 - continuare


4.13 Realizarea elementelor implantabile biocompatibile ale prototipurilor de microsenzori pentru detectia si monitorizarea la distanta a tensiunilor mecanice si temperaturii din implanturi si organe 5.1 Experimentarea si verificarea prototipurilor de microsenzori implantabili in aplicatii preclinice 5.2 Experimentarea si verificarea prototipurilor de microsenzori neimplantabili in aplicatii clinice 5.3 Elaborarea specificatiilor tehnice finale pentru prototipurile de microsenzori realizati

Nota: cu rosu sunt trecute activitatile la care se va renunta.

INCDFT-IFT Iasi - MEDISENS

12

Concluzie Concluzie

Rezultatele obtinute in proiect pana acum, rezultate ce vizeaza materialele utilizate ca elemente de detectie in microsenzori si principiile de functionare ale acestora, ne permit continuarea proiectului si realizarea in ultimul an a microsenzorilor pentru aplicatii medicale propusi initial in proiect, in pofida dificultatilor financiare care au afectat derularea proiectului in primii doi ani.

Multumesc pentru atentia acordata!

INCDFT-IFT Iasi - MEDISENS

13

S-ar putea să vă placă și